                      One-touch 
                        sample analysis, with results in less than 5 secondsRugged 
                        construction for trouble free, reliable measurementsEasy-to-use, 
                        menu driven application softwareConvenient 
                        small size saves laboratory bench spacePerformance 
                        validation routine ensures accurate, repeatable measurement 
                        resultsRS-232C 
                        serial port for direct connection to your PCDiffuse 
                        illumination Spectralon® integrating sphere provides 
                        optimal signal-to-noise performanceMeets 
                        design requirements of the United States Pharmacopeia 
                        23 NF-18 General Chapter 661 containers | 
              
              
              Labsphere’s 
              UV-1000P Ultraviolet Transmittance Analyzer is a convenient, bench-top 
              instrument designed to measure the maximum percent transmittance 
              of "light resistant" pharmaceutical containers. The UV-1000P operates 
              by measuring the diffuse transmittance of the sample in the ultraviolet 
              region (250 to 450 nm). A dual diode array spectrometer with a xenon 
              flashlamp, optimized for UV emission, and an integrating sphere 
              provide instantaneous spectral acquisition. The diffuse illumination 
              geometry of the sphere measures the transmittance from all angles 
              and pathlengths through the sample, and utilizes the total energy 
              from the xenon flashlamp for optimal wavelength stability and flash-to-flash 
              repeatability. The optical design meets the requirements of the 
              United States Pharmacopeia 23 NF-18 General Chapter 661 containers.
              
 The efficiency 
                of the dual diode array spectrometer, coupled with the xenon flashlamp 
                gives reliable and repeatable measurement results in seconds. 
                Instrument software automatically calculates the %T and average 
                values for the UVA and UVB spectral regions and instantly converts 
                the data to give you an accurate rating of the containers ultraviolet 
                blocking ability. Multiple scans can be averaged and viewed simultaneously 
                in easy-to-read formats. ions and instantly converts the data 
                to give you an accurate rating of the containers ultraviolet blocking 
                ability. Multiple scans can be averaged and viewed simultaneously 
                in easy-to-read formats. 
              
              
                 
                  
|  | 
                 
                  |  | 
              
              Windows compatible 
                application software provides an easy-to-use environment with 
                pull-down menus, dialog boxes and built-in report generation functions. 
                The application software guides the operator through analysis 
                without complicated routines. This means minimal amount of operator 
                training. One-touch sample analysis gives results in less than 
                5 seconds and information is quickly converted to a printed report 
                or exported to other spreadsheet or data processing programs for 
                statistical analysis. 
              
              
              A built-in 
                report function generates essential information at the click of 
                a button. Reports include necessary GLP/GMP information such as 
                date, time, operator name, sample identification and test parameters. 
                Reports are viewed on your PC, printed, or exported as text to 
                data processing software for further review. 
               
               
              
              
              
 For validation 
                of the UV-1000P Ultraviolet Transmittance Analyzer’s performance, 
                application software includes an integrated Performance Validation 
                Routine. This allows for on-site validation and revalidation and 
                ensures optimum instrument performance. A set of our four calibrated 
                standards is included with each analyzer to validate the accuracy 
                of both wavelength and photometric scales. The set includes three 
                photometric standards and a wavelength calibration standard. The 
                standards are mounted in specially designed holders that automatically 
                orient and position the standard to ensure precise placement each 
                time the validation routine is performed. The instrument Performance 
                Validation Routine assures quality measurement results as well 
              as satisfying the requirements of regulatory agencies.
  
              
              
                 
                  | Wavelength 
                      Range: 
                   | 250 
                      - 450 nm* 
                   | 
                 
                  | Wavelength 
                      Accuracy: 
                   | ± 
                      2 nm 
                   | 
                 
                  | Transmittance 
                      Accuracy: (nominal +/- uncertainty (%T))
 | 2% 
                      ± 0.5% 
                     10% 
                      ± 1% 
                     20% 
                      ± 1.5% 
                   | 
                 
                  | Data 
                      Interval: 
                   | 1 
                      nm 
                   | 
                 
                  | Effective 
                      Slitwidth: 
                   |  5 nm
 | 
                 
                  | Optical 
                      Geometry: 
                   | (d/0°) 
                   | 
                 
                  | Sample 
                      Beam Diameter: 
                   | 10 
                      nm 
                   | 
                 
                  | Measurement 
                      Range: 
                     TransmittanceAbsorbance
 |  
                      0 - 100%   
                      0 - 2.5A | 
                 
                  | Scan 
                      Time: 
                   | less 
                      than 5 seconds 
                   | 
              
              
               
              * Application 
                software allows for user definable limits within the instrument's 
                wavelength region.
              
              
              
                 
                  | Computer 
                      Interface: 
                   | RS-232C 
                      Serial Port 
                   | 
                 
                  | Data 
                      Transfer: 
                   | 9600 
                      Baud 
                   | 
                 
                  | Min. 
                      Computer Requirements: 
                   | 486/33 
                      processor, Windows 3.1 or higher COLOR VGA 640 x 480 Resolution
 8 MB RAM, 2 MB disk space
 | 
                 
                  | Power 
                      Requirements: 
                   | 110/220 
                      VAC, 60/50 Hz 
                   | 
                 
                  | Operating 
                      Environments: 
                   | 0° 
                      - 500° C, 0% - 70% RH 
                   | 
                 
                  | Dimensions: 
                   | 17H 
                      x 18D x 7W inches 
                   | 
                 
                  | Weight: 
                   | 26 
                      lb. 
                   |