ELITE ISE Na/K/CI or Li
Sodium/Potassium/Chloride or Lithium Analyzer
ELITE Na/K/Cl or Li Specifications:
Sample
Whole Blood, Serum, Plasma, Urine or CSF.
Sample Size
150 µl Whole Blood, Serum or Plasma
150 µl urine (Diluted 3X)
Detection Range
Blood:
Sodium: 60 - 200 mmol/L
Potassium: 1.0 - 10.0 mmol/L
Chloride: 60 - 200 mmol/L
Lithium: 0.1 - 5.0 mmol/L
Urine:
Sodium: 40 - 400 mmol/L
Potassium: 3 - 300 mmol/L
Chloride: 40 - 400 mmol/L
Lithium: (N/A)
CSF
Sodium: (N/A)
Potassium: (N/A)
Chloride: 60 - 200 mmol/L
Lithium: (N/A)
Analysis Time
49 sec (Blood)
50 sec (Urine)
Calibration
Automatic (1-6 hrs.) or On-Demand
Resolution
Blood
Sodium: 0.10 mmol/L
Potassium: 0.01 mmol/L
Chloride: 0.10 mmol/L
Lithium: 0.01 mmol/L
Urine
Sodium: 0.10 mmol/L
Potassium: 0.10 mmol/L
Chloride: 0.10 mmol/L
Lithium: (N/A)
Reproducibility
CV Within Run n=20
Blood, Serum, Plasma:
Sodium: 1.0% or 1.0 mmol/L
Potassium: 1.5% or 0.15 mmol/L
Chloride: 1.5% or 2.0 mmol/L
Lithium: 3.0% or 0.03 mmol/L
Urine:
Sodium: 3.0% or 1.0 mmol/L
Potassium: 3.0% or 0.5 mmol/L
Chloride: 4.0% or 2.0 mmol/L
Lithium: (N/A)
Data Storage
450 patient storage
7 QC Levels and 300 results for each level
Output
40 characters x 4 lines back light LCD display
32 characters per line thermal printer
Serial port (RS-232)
Bar code reader port
Ambient Conditions
Room temperature 15°C ~ 40°C
Power
100-120 ~VAC 50/60Hz, 0.6A
220-240 ~VAC 50/60Hz, 0.35A
Size and Weight
11in/27cm (W) x 16in/41cm (H) x 10in/25cm (D)
26.5 lbs. / 12 kg (without reagent pack)