EasyLyte Na/K/Li
Sodium/Potassium/Lithium Analyzer
EasyLyte Highlights:
- Yes/No prompted operation
- Simple maintenance, and reliability into one economical design.
- The 800mL solutions pack minimizes the need to change reagents and makes cost per test lower than ever before.
- Automatic probe wiping ensures operator safety.
- Calibration is completely automatic and electrodes are maintenance free with long life.
The modular EasySampler connects quickly to the EasyLyte Lithium, enhancing laboratory performance.
|
|
Easylyte Na/K/Li is manufactured by Medica Corporation
EasyLyte Features & Advantages:
The advantages of having an EasyLyte Electrolyte analyzer are numerous.
From Small Hospitals, Intensive Care Units, Cardiac Care units, Medical Office Buildings, Physician's offices to Veterinarians; the EasyLyte performs well under any conditions where electrolyte measurements are required.
Features |
Advantages |
Modular Solution Pack |
Absolutely no handling of waste, which minimizes biohazard contact. Low reagent cost per test. |
Modular Analyzer Design |
Extremely easy to maintain and repair, minimizing downtime of analyzer. |
800 ml Reagent Pack |
Provides the additional cost efficiency of up to 1330 tests per pack. |
Extensive QC & Patient Data Storage |
Convenient access and storage of quality control data. Patient histories immediately accessible for re-evaluation. |
Simple Yes/No Menu Operation |
With simple option changes, all laboratory personnel can be easily trained to operate the analyzer and perform diagnostics. |
Automatic Probe Wiping |
Contact with patient sample is minimized, reducing the need to touch bio-hazardous material. The frequency of probe occlusion is diminished and down time is reduced. |
Automatic Standby Mode |
Reagent consumption minimized, providing cost effective operation. |
EasyLyte Na/K/Li Specifications:
Sample
|
Whole Blood, Serum, Plasma, or Urine
|
Sample Size
|
100 ml Whole Blood, Serum or Plasma
400 ml diluted (1:10) Urine
|
Detection Range
|
Blood:
Na+: 20-200 mmol/L
K+: 0.2-40 mmol/L
Li+: 0.0-5.0 mmol/L
Urine:
Na+: 25-1000 mmol/L (undiluted)
K+: 1.0-500 mmol/L (undiluted)
|
Analysis Time
|
60 sec (Blood)
90 sec (Urine)
|
Data Storage
|
100 patient results
20 normal QC results
20 abnormal QC results
|
Calibration
|
Automatic or On-Demand
|
Resolution
|
Na+: 0.10 mmol/L
K+: 0.01 mmol/L
Li+: 0.01 mmol/L
|
Reproducibility
|
Typical Within Run (n=20)
Blood, Serum, Plasma
Na+: CV < 1% (80-200 mmol/L)
K+: CV < 2% (1-10 mmol/L)
Li+: SD < 0.03 mmol/L (0.2-2.0 mmol/L)
Urine
Na+: CV < 5% (80-200 mmol/L)
K+: CV < 5% (1-10 mmol/L)
|
Output
|
16 place alphanumeric display
40 column printer
Serial port (RS-232)
Auto Sampler port
|
Accuracy
|
Typical vs Flame Photometer
Blood, Serum, Plasma
Na+: r= 0.99 (80-200 mmol/L)
K+: r= 1.00 (1-10 mmol/L)
Li+: r= 0.99(0.2-2.0 mmol/L)
|
Ambient Conditions
|
15-32 C (60-90 F), <85% humidity
|
Power
|
100-115 ~VAC 50/60 Hz, 0.8A or 220 ~VAC 50/60 HZ, 0.4A (Factory set)
|
Size and Weight
|
9.5 in. (24 cm) W x 16.5 in. (42 cm) H x 8.0 in. (20 cm) D, 13 lbs. (5.8 kg)
|
These
materials provided to you by:
Electra Medical Corporation
|