Electra Medical Corporation is a leading provider of Small centrifuges, Angular Rotor Centrifuges, Fixed Angle Rotor Centrifuge, Capillary Tube Centrifuge, basic bench top centrifuges, Stainless Steel Incubators, PID Controlled Incubators, Triple Isolation, incubators, Digital Display Incubators, Cooled Incubators, Air forced ventilation incubators, Medium lab incubators, Natural Convection Dry air sterilizers, PID Controlled Sterilizers, Stainless steel ovens, Vacuum ovens, Vacuum Drying Function autoclave, bench top autoclaves, Various capacities autoclaves, small steam sterilizers, large capacity steam sterilizer, unstirred water baths, general purpose water baths, shaking water bath, circulator/water baths
Blood Gas Analyzer

EasyBloodGas Specifications

CLIA Classification: Moderate complexity
Sample Type: Whole blood
Sample Size: 100 µL Syringe 75 µL Capillary
Measured Parameters
pH 6.900 - 7.900 pH units
PCO2 8.0 - 150.0 mmHg
P02 10-700 mmHg
Calculated Parameters
pH (T)
(pH temperature corrected)
PCO2 (T)
(PCO2 temperature corrected)
P02 (T)
(P02 temperature corrected)
TCO2
(Total Carbon dioxide)
0 - 50 mmol/L
HC03
(Bicarbonate)
0 - 50 mmol/L
BEb
(Base Excess in blood)
-25.0 to 25.0 mmol/L
BEecf
(Base Excess in extra cellular fluid)
-25.0 to 25.0 mmol/L
SBC
(Standard Bicarbonate)
0 - 50 mmol/L
%S02c
(Oxygen Saturation at normal P50)
40.0 - 100.0%
A-aDO2
(Alveolar arterial oxygen gradient)
0 - 700 mmHg
RI
(Respiratory Index)
0.0 - 70.0
Input Parameters
Patient Temperature 5 - 45°C
Hemoglobin 3.0 - 30.0 g/dL
F102 10-100%
Patient ID 14 digits
Sample Temperature Control: 37°C ± 0.2°C
Ambient Conditions: 15 - 30°C (59-86°F), 500 - 800 mmHg (max 15 PSI) 5 - 85% relative humidity, non-condensing atmospheric air environment (21% P02)
Analysis Time: 125 seconds
Data Storage: 64 Patient results with Operator ID, Patient ID, Date and Time QC - up to 30 results for each Level (1, 2, 3)
Calibration: Automatic or On-Demand
Input/Output: Numeric keypad, graphic display, 27 column thermal printer, barcode reader port, RS-232 computer interface port
Power: 100/115-VAC, 50-60 Hz, 0.8 A or 220—VAC, 50-60 Hz, 0.4 A
Size & Weight: 14.5W x 1 2.5”H x 7”D (37cmW x 32cmH x 1 8cmD), l6Ibs (7.3kg) with reagent module

Manufactured by Medica Corporation